Servo Service ManualE382 E407E 061 01 01 02ImportantServoNotes1-2Siemens-Elema ABE382 E407E 061 01 01 02ServoImportant ContentsContents 1. Description of functions.34. Disassembling and assembling.45. Service procedures.56. Preventive maintenance.78. Diagrams.9E382 E407E 061 01 01 02Siemens-Elema AB1-3ImportantServoImportant1GeneralSymbols used in this manual.

SERVO ventilators have been under continuous development since their introduction in 1971 40 years Innovation leaders The SERVO family exceeds 100.000 units worldwide 100.000Used worldwide SERVO-U® delivers many effective options for protective ventilation. All of them more accessible, understandable and easy to implement. The Servo-i Ventilator System be performed as part of a service contract with MAQUET. 1.1.7 Disclaimers.Improper Use MAQUET has no responsibility for the safe operation of Servo-i Ventilator System if the Intended Use requirements specified in this document are not followed. SVX-128EN User Interface Patient Patient Unit breathing system.

Service documentation for the Servo ™ Ventilator System consists of: – User's manual – Service Manual – Spare Parts List. ESD sensitive components.

Make sure to take precautions to avoid damaging ESD sensitive components. Documentation for all optional equipment included in the Servo System is also available. The User's manual is an indispensable complement to the Service Manual for proper servicing. Serial number of the unit is found on a label attached to the Patient Unit close to the supply gas inlets. System version number can be found in the Status window on the User Interface. Make sure that the version of the User's manual corresponds to this system version.

Text inside a box is used to highlight important information. In addition to the Important information given here and in the related documents (e. In the User's manual), always pay attention to applicable local and national regulations. Responsibility for the safe functioning of the equipment reverts to the owner or user in all cases in which service or repair has been done by a non-professional or by persons who are not employed by or authorized by Siemens, and when the equipment is used for other than its intended purpose. Data on internal pressures in the Servo System are given in Pa (bar). Airway pressures are given in cm H2O (Pa).

1 hPa = 1 mbar 1 mbar = 1 hPa 1 kPa = 10 mbar 1 mbar = 0.1 kPa 1 kPa = 0.01 bar 1 bar = 100 kPa 1 kPa N 10 cm H2O 1 cm H2O N 0.1 kPa 1 kPa N 0.01 at 1 at N 100 kPa 1 kPa N 0.01 kgf/cm2 1 kgf/cm2 N 100 kPa 1 kPa N 0.01 kp/cm2 1 kp/cm2 N 100 kPa 1 kPa N 0.145 psi 1 psi N 6.9 kPa1-4. Special waste. Make sure to discard worn-out batteries and other disposable parts according to local regulations and in an environmentally acceptable way. Recycle if possible. Recycling facilities may not be available in all areas. Technical training. Refers to the Servo Technical training supplied by Siemens.

Service contract. Refers to the Servo Service contract supplied by Siemens.Hazard notices. Before disassembling or assembling the Servo System, make sure that the: – Gas supply is disconnected. Crack corel x5 remove protexis inc. – Mains power cable is disconnected. – On/Off switch is set to Off. – Battery modules are disconnected.

– The Servo System is cleaned according to instructions in the User's manual, chapter Routine cleaning and chapter Regular maintenance, section Extended cleaning of insp. With power supply connected to the Servo System, there are energized electrical components inside the unit. All personnel must exercise extreme caution if fault tracing or adjustments are performed with power supply connected and with user interface and patient unit covers removed.Siemens-Elema ABE382 E407E 061 01 01 02ServoImportantImportant InstallationTo the responsible service personnel. Only personnel trained and authorized by Siemens shall be permitted to install the Servo System.

Servo

The installation and handing over procedures are described in the ”Servo Ventilator System – Installation Instructions”. The contents of this document are not binding. If any significant difference is found between the product and this document, please contact Siemens for further information. We reserve the right to modify products without amending this document or advising the user.Functional check. After any installation, maintenance or service intervention in the Servo System, perform a Pre-use check according to instructions in the Servo Ventilator System – User's manual.Service. The Servo System must be serviced at regular intervals by personnel trained and authorized by Siemens. Any maintenance or service must be noted in a log book provided.

Only personnel trained and authorized by Siemens shall be permitted to perform installation, service or maintenance of the Servo System. Steam folder location. Only Siemens genuine spare parts must be used. PC boards (spare parts) must always be kept in a package for sensitive electronic devices. Siemens will not otherwise assume responsibility for the materials used, the work performed or any possible consequences of same. The device complies to standards and requirements as stated in the Servo Ventilator System – User's manual. It is recommended that maintenance and service is done as a part of a service contract with Siemens.

Preventive maintenance must be performed at least once every year as long as the unit is not used more than normal. Normal operation is estimated to correspond to approx. 5.000 hours of operation.

Maquet Servo 300 Manual

Details are found in this Service Manual, chapter ”Preventive maintenance”. Worn-out batteries must be recycled or disposed of properly according to local regulations. Recycle facilities may not be available in all areas. Batteries must not be disposed of with ordinary waste.

All other disposable parts shall be discarded according to local regulations and in an environmentally acceptable way. When working with ESD sensitive components, always use a grounded wrist band and a grounded work surface. Adequate service tools must always be used.E382 E407E 061 01 01 02Siemens-Elema AB1-51ImportantServoImportant1Environmental declarationConstruction materialsPurposeThe construction materials used in Servo in% of the total weight.This environmental declaration is for a Servo basic unit including the carrier and one battery. Letters codes within brackets refers to the Functional Block Diagram in chapter Diagrams.